Calibration Services
Resistance Standards
Temperature Standards
Humidity Standards
Measuring Instruments
Analytical Instruments
Optical Instruments
Force Standards
Pressure Standards

Dead Weight Force and Load Cell
Calibrating Machines



The Ultimate calibration Standard



Description


Of all the devices and machines designed to measure mechanical forces none has ever equaled the positive and unfailing accuracy obtainable with dead weights.

Morehouse has two different models of Dead Weight Load Cell Calibrators available: The Model L and the Model MH.

The Model L is available only in capacities of 1,000 lbf or less – standard capacities being 500 lbf and 1,000 lbf, or the approximate equivalent in kilograms or Newtons: 250 and 500 kgf respectively, or 2,500 and 5,000 Newtons respectively.

The Model MH is available in larger capacities – standard capacities being 2,000 lbf, 5,000 lbf, 10,000 to 100,000 lbf, or the approximate equivalent in kilograms or Newtons.


Construction

Both models of the machine are designed to make full use of the accuracy of dead weights without any mechanical interference or machine losses. The yoke assembly on which the dead weights are suspended bears directly on the instrument being calibrated.

There are no intervening levers or flexures between the weight complements of the machine and the instrument being calibrated to introduce ratio and beam deflection errors.

The loading stages on both models are adjustable to accept instruments of varying heights and to allow for tension as well as compression calibrations.

The weights supplied with both machines are machined from stainless steel and are adjusted to better than .003% of their nominal weight. The accuracy of the weights can be certified traceable to the National Institute of Standards and Technology.


Specifications

Capacities Available:
MODEL L : 500 lbf, 250 kg, and 2,500 N
  : 1,000 lbf, 5,000 kgf, and 5,000 N

MODEL MH
: 2,000 lbf – 120,000 lbf and the approximate equivalent in kilograms and Newtons. Special capacities within the weight limitations and special weight complements are available in both models.


CALIBRATION MODE:
:Tension and Compression on both Models

WEIGHTS  
Material
: Stainless Steel
MODEL L
: Weights are applied and removed through a system of manually Operated levers
MODEL MH
:Weights are applied and removed pneumatically with air Controlled selector switches. Operating air pressure: 10 p.s.i.
ACCURACY
:All weights and yoke assembly are calibrated in force or mass to within 0 .003%. Certified traceable to the National Institute of Standards and Technology.


Morehouse 1000 lbf
Model L
Dead Weight Force Calibrator

The Morehouse Model MH Dead Weight Load Cell Calibrator

 
 
 
Dead Weight Load Cell Calibrators
Model L and Model MH
 
Model MH General Information
STANDARD YOKE DIMENSIONS
The dimensions of the machine’s yoke assembly may be varied to accommodate individual requirements with respect to overall dimensions of load cells or other instruments to be calibrated in both tension and compression.
 
 
 


CopyRight © 2006 Vaiseshika Electron Devices