Load Cells |
|
Morehouse Ultra Precision Calibration Grade Load Cells
|
Morehouse Universal Calibration grade load cells from 500 LBF to 100,000 LBF.
ASTM-E74 Class AA lower limit around 10% of range in compression and tension when calibrated in Mv/V or combined with the morehouse load cell meter model 4215 or comparable meter.
ASTM-E74 Class A lower Limit around 2% of range in compression and tension.
Uncertainties of around 0.005% of capacity in compression and tension when combined with the morehouse load cell meter model 4215.
NOTE: Morehouse Calibration Grade Load Cells above 50,000 LBF will have a guaranteed uncertainty of around .01% of capacity or better (Class AA lower limit of better than 20%). Morehouse is the only accredited commercial laboratory that can calibrate class AA lower limits up to 120,000 LBF. |
|
Morehouse Precision Calibration Grade Load Cells
|
|
Morehouse High Precision Universal Load Cells from 500 LBF to 200,000 LBF.
Guaranteed ASTM-E74 Class A lower limit better than 4 %of range in compression and tension when combined with the morehouse load cell meter model 4215.
NOTE: Morehouse Precision Calibration Grade Load Cells above 100,000 LBF will have a guaranteed uncertainty better than .025% of capacity. |
Morehouse General Purpose Calibration Load Cells
|
Morehouse Load Cells from 500 LBF to 200,000 LBF. First choice for structural and materials testing.
Guaranteed ASTM-E74 Class A lower limit better than 8% of range in compression and tension when combined with the morehouse load cell meter model 4215 or comparable meter. |
|
NOTE: We also sell higher capacity load cells, dual bridge load cells and s-beam load cells and will Guarantee a Class A lower limit of better than 10% of capacity from 300 LBF to 1,000,000 LBF when combined with the morehouse load cell meter model 4215. |
Morehouse Load Cells are Black in color. Custom painting is available per additional charge if we do not stock the desired color paint. |
Morehouse Load Cell Meters |
 |
| |
Load Cell meter with Fully Bipolar Range +/-999,999 Display, 24 bit resolution, Auto Setup for 25 Load Cells
Standard Features:
| |
Fully Bipolar Range +/-999,999 Display
|
| |
24-Bit Resolution
|
| |
Non-linearity 0.005%
|
| |
2-line X 20 Character Vacuum Fluorescent Display
|
| |
Auto Load Cell Identification
|
| |
User Input Text Label on Second Line of Display
|
| |
Auto Setup for 25 Load Cells
|
| |
Front Panel Shunt Calibration with Two Selectable Resistors
|
| |
Functions Included: Peak / Valley / Hold / Tracking / Peak Reset
|
| |
Quad Limits Output: Independent, Isolated Solid State Relays, Controls AC or DC Signals
|
| |
Six User-Selectable Filters
|
| |
Tri-state Limit Display: On, Off, Disabled
|
| |
Automatic Display Unit Conversion: Lb, Kg, N Psi, Mpa, Klb, Kn, t
|
| |
D/A Output: 10V std; 5V
|
| |
Direct Analog Amplifier Output
|
| |
RS485 std.
|
| |
Full Operation and Calibration via RS232
|
| |
5-Point Linearization
|
| |
OPTIONAL FEATURE (pictured above) - MOREHOUSE SOFTWARE TO ALLOW MODEL 4215 TO BECOME DIRECT READING UNIT IF CALIBRATED IN mV/V (This requires the user to have a laptop with RS232 or USB port and uses coefficients generated from the calibration report to read in direct engineering units of LBF, KGF, or N)
|
|
| |
|
|