Calibration Services
Resistance Standards
Temperature Standards
Humidity Standards
Measuring Instruments
Analytical Instruments
Optical Instruments
Force Standards
Pressure Standards
 

"Vaiseshika" Digital Thermocouple Test Set and Calibrator
Type : 7709 EP

 

Description
Digital Thermocouple Test Set and Calibrator type 7709 EP is an EPROM (Erasable Programmable Read-only-Memory) based design. The instrument can accept J, K, S, R and T type of thermocouples in the same equipment. The instrument is ideal to measure temperatures with J, K, S, R and T type of thermocouples as well as for the calibration of temperature indicators employing J, K, S, R and T type of thermocouples. excellent liberalization of the order of 1șC is attained over the complete range of thermocouple viz. J, K, S, R, and T.

4 Digits Digital Thermocouple Test Set and Calibrator is a versatile equipment in the area of thermal engineering. The instrument is ideal to calibrate temperature gauges, potentiometric recorders, platinum resistance thermometers, temperature controllers and recorders. The instrument can also be used as a thermocouple simulator. A calibrated stable potential source, continuous variable from 0 to ± 100 mV, is provided with the test set. The potential variation can be done with a resolution of 1șC. Similarly temperature sensed by a thermocouple in a furnace can be monitored with Digital Read Out Display in this instrument.

A standard Wheatstone bridge is also provided with the test set. The resistance ranging from 0.1 ohm to 1K ohm can be measured and simulated with an accuracy of ± 0.1%.

Features
  • Reliable electronic design.
  • User friendly calibrators.
  • Excellent calibration facilities with digital read out.
  • Resistance calibration facility from 0.1 ohm to 1K ohm for RTDs and thermistors.
  • Complete coverage to J, K, S, R, and T type of thermocouples and temperature gauges.
Specification
Temperature   measurement : Ambient to 750șC for 'J' type
: Ambient to 1350șC for 'K' type
: Ambient to 1700șC for 'S' type
: Ambient to 1700șC for 'R' type
: Ambient to 400șC for 'T' type
Calibration   facility : For the calibration of
: J, K, S, R, and T type of temperature indicators.
Simulation   facility : 0 to ± 100 mV for any thermocouple with digital read out in
 corresponding șC.
Accuracy : ± 0.3% of mV for T, J & K type thermocouples, ± 0.2% of mV
 for R & S type   thermocouples
Resolution    for   calibration : 1șC
Ambient   temperature compensation : Automatic
Digital   display   during calibration : in Degree Celcius
Power : 230 V ± 10% AC, 50 Hz

Capability

EPROM based Digital Thermocouple Test Set and Calibrator Type 7709 EP can function as a Digital Temperature Indicator as well as Calibrator for any of the J, K, S, R, and T type of Thermocouples. The choice of thermocouple is done through a manual selector switch.
Wheatstone Bridge facility :  (for the calibration of RTDs / Pt100 sensors and resistane   measurements)
Resistance Range :  0.1 ohm to 1K ohm
Resolution :  0.1 ohm
Accuracy :  ± 0.1%
Switch configuration :  Brass switches with silver plating
Housing :  Table top metallic cabinet
Dimensions :  48(L) x 35(B) x 13(H) cm.
Weight :  7.5 Kg.


CopyRight © 2006 Vaiseshika Electron Devices