2. Calibration Grade Load Cells

Products

Morehouse Load Cell Meter

Description

Load Cell meter with Fully Bipolar Range +/-999,999 Display, 24 bit resolution, Auto Setup for 25 Load Cells 

 

Standard Features:
Fully Bipolar Range +/-999,999 Display
24-Bit Resolution
Non-linearity 0.005%
2-line X 20 Character Vacuum Fluorescent Display
Auto Load Cell Identification
User Input Text Label on Second Line of Display
Auto Setup for 25 Load Cells
Front Panel Shunt Calibration with Two Selectable Resistors
Functions Included: Peak / Valley / Hold / Tracking / Peak Reset
Quad Limits Output: Independent, Isolated Solid State Relays, Controls AC or DC Signals
Six User-Selectable Filters
Tri-state Limit Display: On, Off, Disabled
Automatic Display Unit Conversion: Lb, Kg, N Psi, Mpa, Klb, Kn, t
D/A Output: 10V std; 5V 
Direct Analog Amplifier Output
RS485 std.
Full Operation and Calibration via RS232
5-Point Linearization
Optional Feature (pictured above) - Morehouse software to allow Model 4215 to become Direct Reading Unit if Calibrated in mV/V (This requires the user to have a laptop with RS232 or USB port and uses coefficients generated from the calibration report to read in direct engineering units of LBF, KGF, or N)